Main

Tissue oxygenation is a critical determinant of organ function. In clinical settings, tracking oxygenation is limited to indirect approaches with suboptimal utility. In the context of organ transplantation1, vascular complications, such as hepatic artery thrombosis following orthotopic liver transplantation, frequently lead to graft ischemia, dysfunction and loss2. The ability to continuously monitor graft oxygenation status via a wireless sensor placed at the time of transplant would augment our abilities to track tissue perfusion, allowing timely intervention, and would help differentiate parenchymal rejection from graft vascular disease in the setting of graft dysfunction.

Most clinically available direct O2 monitoring systems rely on electrochemical sensors (Clark-type electrodes)3 and luminescence probes4. Clark-type electrodes consume O2, which potentially confounds accurate measurement of tissue O2, especially at low O2 concentrations, and are not well suited for long-term measurements due to their susceptibility to biofouling3. This led to the development of luminescence O2 probes. These probes, although easy to use and accurate, are invasive and fragile and currently rely on wired connections, limiting their use in the hospital environment. Fluorine magnetic resonance imaging5 and electron paramagnetic resonance spectroscopy6 have also been used to directly probe local O2 content in vivo. However, these methods suffer from an inability to provide real-time data, as they require long scanning times to quantify O2 concentration with sufficient resolution.

As an alternative to direct O2-sensing approaches, near-infrared spectroscopy7,8 has demonstrated the ability to quantify tissue O2 saturation levels. However, the use of noninvasive near-infrared spectroscopy systems, especially in deep tissues, is limited by the penetration depth of near-infrared light. An implantable wireless oximeter has demonstrated regional O2 saturation measurements from just below the skin in an untethered, freely moving mouse in a cage; the system relied on near-field electromagnetic (EM) waves for wireless power transfer, restricting it to shallow operation9.

Ultrasound (US) linkage has recently been explored as an attractive option for millimeter-scale deep-tissue implants10,11. In contrast to EM waves, US allows for wireless power and data transfer via pressure waves with millimeter and submillimeter wavelengths (1–10-MHz US carrier frequency). As a comparison, 2-MHz US waves have a wavelength of 0.75 mm, while 2-GHz EM waves have a wavelength of 25 mm; this enables efficient coupling to miniaturized implants. In addition, US experiences lower attenuation in soft tissues (for example, US attenuates ~1–2 dB cm−1 at 2 MHz; EM attenuates ~10–12 dB cm−1 at 2 GHz)12 and consequently has a higher FDA regulatory limit for power flux density (US, 720 mW cm–2; EM, ~10 mW cm–2)13,14. Unlike that of EM waves, US power transfer through bone is more challenging because US attenuation in bone is more than an order of magnitude higher than in soft tissue, and reflections and scattering occur at the soft tissue–bone interface15.

In recent years, wireless systems based on ultrasonic millimeter-sized implants have demonstrated electrical recording10 from and stimulation11 of the sciatic nerve in an anesthetized rat model and tumor oxygenation16 and photodynamic tumor therapy17 in an anesthetized mouse model. Here, we present an ultrasonic wireless implant system for continuous real-time O2 measurements at centimeter-scale depths demonstrated in a clinically relevant anesthetized sheep model, showing promise of avoiding the drawbacks of the current O2-sensing technologies mentioned above. The system uses an O2 sensor implant that incorporates a single piezoelectric ceramic and a luminescence sensor consisting of a µLED, an O2-sensing film, an optical filter and a state-of-the-art integrated circuit (IC)18. The sensor operates on the same readout principle, phase luminometry, as most luminescence sensors19 and probes20, but with much lower power consumption and competitive or better O2 resolution than these sensors and probes. This is achieved by compact integration of the sensor components and a custom IC. Wireless power and bidirectional data transfer are enabled using a single US link from an external transceiver. The use of a single link protocol combined with the tight integration of wireless sensor components results in a very small sensor volume (4.5 mm3). The small volume and lightweight construction (~17.4 mg) provide potential capabilities for use with minimal tissue damage. The system operation is demonstrated in vitro in distilled water, PBS and human serum, ex vivo in porcine tissue and in vivo in sheep.

Results

A custom integrated circuit and a single ultrasonic link enable a complete wireless optical sensor in a volume of 4.5 mm3

An ultrasonic transceiver directs US energy from outside of the body toward an implant placed in muscle or deep tissue (Fig. 1a). This wireless acoustic link provides power to the sensor and allowed bidirectional data transmission. A custom mixed signal system drives the external transducer. The implant consists of a piezoelectric crystal for energy collection and communication coupled to a luminescence sensor for O2 detection.

Fig. 1: Wireless O2-monitoring system overview.
figure 1

a, Schematic of the system as demonstrated in this paper for local tissue O2 monitoring in sheep. The free-floating wireless O2 implant is surgically placed beneath the biceps femoris muscle, and the wound is closed with sutures. The external US transducer, placed on top of the closed surgical site, establishes a wireless acoustic link to the implant. The external transducer powers the implanted sensor by delivering acoustic energy through the tissue and then listens for backscatter reflections from the sensor’s piezoelectric crystal, in which O2 data are encoded. The external transducer is driven by a custom mixed signal system, including decoding and storage of the wireless O2 data received by the external transducer. b, An expanded view of the wireless sensor platform, including a lead zirconate titanate (PZT) piezocrystal and a luminescence sensor, consisting of a µLED, a 3D-printed µLED holder, an O2-sensing film, an optical filter and an IC. c, The following steps were used for wireless sensor fabrication: (1) a piezocrystal was bonded with conductive silver epoxy, wire-bonded to a flexible printed circuit board (PCB) and encapsulated with parylene-C; (2) other sensor components were assembled on the flex board, and only the regions where the wire bonds are located were encapsulated with ultraviolet (UV)-curable epoxy; (3) the small gap (~50 µm) between the film and the µLED holder was filled by PDMS and (4) sensor parts excluding the piezocrystal were encapsulated with highly O2-permeable black silicone. d,e, Schematic (d) and photograph of the wireless sensor before black silicone encapsulation and size comparison to a finger (e); scale bar, 5 mm. f, Top view of the 4.5-mm3 packaged wireless sensor with the piezocrystal at the bottom, the luminescence sensor at the top and a size comparison to a United States dime.

The entire implant is manufactured with standard medical implant processes

The implant (Fig. 1b) consists of a piezoelectric crystal (750 × 750 × 750 µm3), a µLED, an O2-sensing film, an optical filter, an IC and a holder providing mechanical support during wire bonding of the µLED. To assemble the sensor, a piezocrystal is attached and then encapsulated with parylene-C; the remaining components are then assembled on the flex board (Fig. 1c). Wire bonds are encapsulated with UV-curable epoxy, which was chosen due to its high bond strength. Finally, the sensor part is coated with black silicone to avoid background interference by the luminescence of tissue or blood21. The implant (Fig. 1d–f) measures 3 × 4.5 × 1.2 mm3, occupies a volume of 4.5 ± 0.5 mm3 and has a detection volume of ~0.26 mm3. The piezo resonant frequency determines the US link carrier frequency, as this frequency is set by the crystal thickness and aspect ratio22; crystal geometry was chosen to maintain a reasonable trade-off between the frequency-dependent acoustic loss in tissue, the power harvesting capacity and the impact on total implant size.

A complete biocompatible phase luminometric O2 optrode is built into the implant

The biocompatible O2-sensing film (Fig. 2a) consists of polydimethylsiloxane (PDMS) containing silica particles with surface-adsorbed ruthenium dyes23. Upon excitation at ~465 nm, the ruthenium dyes emit light with a peak intensity at ~621 nm (Fig. 2b). The excited ruthenium dyes undergo collisional quenching with O2 molecules, leading to a reduction in luminescence intensity and lifetime21. Both the intensity (Ι) and lifetime (τ) depend on O2 concentration according to the Stern–Volmer equation19: I0/I = τ0/τ = 1 + KSV[O2], where I0 and τ0 (~6.4 µs) are the intensity and lifetime at an O2 concentration of 0, respectively, and KSV is the Stern–Volmer constant. Either intensity or lifetime can be measured to compute O2; lifetime is independent of variations in light source intensity, dye concentration, inner filter effects and photobleaching (to a large extent), all of which are limitations of intensity-based sensors24,25.

Fig. 2: Biocompatible O2-sensing film, operating principle of the luminescence O2 sensor and its optical characterization.
figure 2

a, An expanded cross-sectional view of the O2 sensor. The inset illustrates a model of the locus of ruthenium dye and O2 molecules in silica-containing PDMS. The orange squares and black circles represent the ruthenium dye and O2 molecules, respectively. The dye-adsorbed silica particles were dispersed in PDMS. b, Normalized absorption and emission spectra of the ruthenium dye in the O2-sensing film, along with the normalized emission spectrum of the blue µLED and transmission spectrum of the optical filter. a,b, In operation, the µLED, placed on the sensor platform to illuminate the film, produces light with a peak intensity at ~465 nm that excites the ruthenium dye. The excited dye emits luminescence with a peak at ~621 nm. An optical filter suppresses excitation light and transmits luminescence, enabling emission detection by an IC with an integrated photodiode. c, Illustration of frequency domain excitation and emission signals. d, Current–voltage–light output characteristics of the blue µLED (n = 1). e, Responsivity spectrum of an integrated, nwell/psub photodiode with an active area of 300 × 300 µm2 and a 0.6-V reverse-bias voltage. The orange-shaded area highlights wavelength ranges with strong luminescence emission. f, Photobleaching of ruthenium dye in the O2-sensing film under continuous square-wave illumination with a peak excitation light power of ~1.53 µW at room temperature in air (21% O2) for a 60-h period (n = 1 sensor). g, Normalized luminescence intensity of ruthenium dye versus time after immersion of the same O2 sensor (n = 1) used for the photobleaching test in PBS solution at 37 °C in room air. During the test, the sensor was operated with the same operating conditions as in the photobleaching test. The continuously collected data were averaged every 3 h (f) and 12 h (g) and are presented as the mean ± 1 s.d. for each data point.

This sensor operates on the principle of phase luminometry, wherein the phase shift (ΔФ) between the excitation and emission light is measured to determine the luminescence lifetime. During operation, the emitted light is square-wave modulated at a fixed frequency (fop = 20 kHz) (Fig. 2c), exciting the ruthenium dye with a peak excitation power of ~1.53 µW, resulting in an average power intensity of ~4.9 µW mm–2 at the film surface (Fig. 2d). At 37 °C in room air, the ruthenium dye produced an emission with a typical average power density of ~8 nW mm–2 at the same fop, but with a ΔФ relative to the phase of the excitation light. After the excitation light is suppressed by an optical filter, emission is detected by a reverse-biased, on-chip photodiode with a responsivity of ~0.12 A/W at ~621 nm peak emission (Fig. 2e). The resulting ΔФ = tan−1(2πfopτ) ≈ ωopτ for ωopτ 1 and is dependent on the luminescence lifetime that, in turn, is related to O2 via the Stern–Volmer equation.

Ruthenium dye photobleaching was evaluated using a fully packaged sensor continuously operated in room air and at room temperature for a 60-h period. The luminescence intensity decreased rapidly to ~96.8% of its initial value in the first 10 h and at a slower rate from ~96.8% to ~93.6% in the next 50 h (Fig. 2f), indicating that the sensor could be operated with 1% duty cycle, corresponding to 14.4 min of continuous operation in a single day, for 250 d with an only ~6.4% intensity drop. After the bleaching test, a long-term test of the same sensor immersed in 1× PBS at 37 °C was conducted in room air to evaluate dye leaching; fluctuation in luminescence intensity was within ±1.7% and did not exhibit a decreasing trend in 14 d (Fig. 2g).

A custom IC drives sensor operation and communication for <150 µW average power consumption

A system block diagram is shown in Fig. 3 (see ref. 18 for details). During operation, the controller driving the external transducer alternates between transmit (TX) and receive (RX) modes. When in TX mode, it drives the transducer with a 2-MHz carrier wave onto which it encodes digital information in discrete pulses. The implant begins harvesting energy following the arrival of an US pulse; this energy is rectified and stored on the on-chip capacitor. The stored energy allows the implant to operate when the external transducer is in RX mode. The rectified voltage (Vrect) is regulated at 1.2 V. When the regulator voltage exceeds a predetermined level (~1 V) during the power-up period, a power-on-reset (POR) signal is triggered to initialize the IC. After an initialization period (Supplementary Fig. 1), an LED driver drives the µLED with a 24-µA, 20-kHz square-wave current, starting the O2-sensing operation. The µLED excites the O2-sensitive dyes, producing luminescence. The light induces a photocurrent (IPD) on the photodiode, which is converted to a voltage. The resulting AC voltage is compared to its DC component by an analog front end (AFE) for zero-crossing detection26, generating a time-delayed signal. A time-to-digital converter converts the time delay (phase difference, ΔФ) between the reference signal (Фref) used to drive the µLED and the luminescence signal (ФPD) into 10-bit digital data. The 10-bit data are serialized and divided into two equal 15-µs-long data packets with a preamble and a postamble by a finite-state machine; the first packet contains the most significant bits (MSBs). The measured minimum detectable average optical power, yielding a signal-to-noise ratio (SNR) of 1 at a 1-Hz bandwidth, was ~1.3 pW at ~621 nm peak emission for the optical readout, dominated by the AFE noise. The SNR was ~53 dB under a typical ~6.7-nW mm–2 light power after the optical filter for the sensor, operated in room air at 37 °C.

Fig. 3: Entirely wireless O2-monitoring system block diagram.
figure 3

a, The external transceiver is shown in the top left, the US link is shown in the top middle and the wireless sensor is shown on the right. Top left, the external transceiver consists of TX and RX paths, where the TX path encodes downlink data onto a 2-MHz carrier. During TX operation, a level shifter boosts a low-voltage TX signal from a digital controller, and a high-voltage pulser drives an external piezo transducer. The RX path is enabled when the TX path is disabled. The reflected US backscatter from the sensor’s piezocrystal is captured by the same external piezo transducer and is digitized by the RX chain. Top middle, the external piezo coupled to the outside surface of tissue produces US waves traveling through tissue; these arrive at the sensor after one time-of-flight (ToF). The downlink provides power and a transmit command for the sensor. The uplink consists of amplitude-modulated backscattered US waves that arrive at the external piezo 2ToF after being sent during TX. Right, the sensor IC architecture; CLK, clock. b, Timing diagram. The IC rectifies electrical power collected by the sensor’s piezo, and the POR initializes the O2-sensing operation. The on–off keying (OOK) demodulator detects the downlink US envelope, producing a notch. The first notch starts uplink transmission. Two data packets are transmitted via digital amplitude modulation of the US backscatter; the first packet contains five MSBs and a 1-bit preamble. The uplink data are received by the external piezo when the external transceiver is switched to RX. The uplink transmission stops when the third notch duration is longer than ~64 µs (127 oscillations of a 2-MHz carrier). The IC circuitry is duty cycled during uplink transmission to reduce energy consumption.

An uplink data transmission begins when the demodulator detects a ‘falling edge’ in the US data input from the external transceiver, generating a notch (VOOK). The notch serves as a reference to time synchronize the sensor IC and the external transceiver during uplink transmission; a data packet is transmitted after a notch. Data packets are encoded in the US reflections from the sensor’s piezo and transmitted via backscatter amplitude modulation. Backscatter amplitude modulation is achieved by modulating the electrical load resistance (Rload), in shunt with the piezo impedance (Zp), through a modulation (transistor) switch controlled by Фmod, changing the US reflection coefficient at the piezo boundary (Supplementary Fig. 2) and thus the backscatter amplitude27. When the switch is turned on to transmit the O2 data, the Rload is reduced from a resistance higher than 80 kΩ, depending on the IC power consumption and the rectifier input voltage amplitude, to ~0.5 kΩ (switch-on resistance). The uplink transmission stops when the notch duration is longer than ~64 µs.

The total area of the IC die is ~3.84 mm2 (Supplementary Fig. 3). The minimum electrical power required for proper IC operation is ~150 µW during the O2-sensing phase. The AFE, LED driver, voltage doubler and time-to-digital converter are duty cycled off during uplink transmission, reducing IC power consumption to ~22 µW and avoiding the need for a large off-chip Cstore. The average power dissipation of the IC drops to less than 150 µW, including the rectifier’s power conversion efficiency during operation, depending on the O2 sampling rate (fs). The system fs is externally controlled through the external transceiver.

During operation, the transceiver is switched from TX to RX mode to capture uplink data. The RX path decodes the backscatter, generating real-time O2 data. To avoid overlapping TX and RX pulses, the data packet duration (TDM) is kept shorter than the round-trip ToF (2ToF) of an US pulse between the sensor’s piezo and the external transducer, limiting the minimum operating depth. To overcome this limitation, the digital data are divided into two parts that are subsequently transmitted. See also Supplementary Fig. 4 for an alternative communication protocol for sensors implanted deeper than 5 cm.

The system samples O2 at 350 Hz with a bit error rate of <10−5 and a resolution of <5.8 mm Hg Hz–1/2 across the physiologically relevant O2 range (0–100 mm Hg)

The system was first characterized in a water tank (Fig. 4a); water has an acoustic impedance similar to that of soft tissue (~1.5 Mrayl)28. Measured waveforms and backscatter signal were recorded for the system operated at a depth of 5 cm and a fs of 350 Hz using an acoustic field with a spatial-peak time-average intensity (ISPTA) of 237 mW cm–2 (~32.9% of the FDA safety limit; ISPTA derated is 720 mW cm–2 for diagnostic US) (Fig. 4b). The system exhibited a power transfer efficiency (PTE; Supplementary Notes) from the acoustic power at the implant’s piezo surface to the IC’s electrical input power of ~20.4% and an US link PTE of ~3.9%, defined as the ratio of the IC’s input power to the acoustic power emitted from the external transducer. Each data packet was 15 µs long (Fig. 4c); the first packet has a ‘1’ preamble and 5-bit MSBs, and the second packet has 5-bit least significant bits (LSBs) and a ‘1’ postamble. The system achieved a modulation depth (MD) of ~41% and an uplink bit error rate (BER) < 10−5 (0 of 121,000 samples) with the optimal threshold, minimizing BER (Fig. 4d).

Fig. 4: In vitro characterization of the wireless O2-sensing system.
figure 4

a, Wireless sensor operation in distilled water at 37 ± 0.1 °C (physiological temperature). The depth was set to 5 cm, and water O2 levels were precisely controlled using two identical gas flow controllers and a commercial O2 probe (NEOFOX) during measurements. b, Sensor waveforms and backscatter signal were recorded in the wireless measurement of a single O2 sample. c, Zoomed-in data packets show digital amplitude modulation of the US backscatter. d, Backscatter relative difference for 121,000 O2 samples, showing a ~41% MD. To detect each bit, which is either ‘0’ or ‘1’, the amplitude of the backscatter signal in the data packets is compared to a threshold. The wireless system achieved an uplink BER < 10−5 (0 of 121,000 samples) in the measurement, demonstrating a robust data uplink. e, System response to various O2 concentrations. The measurement was performed at 350 Hz fs and a 5-cm depth in the O2 range of 3.8–150 mm Hg. f, Allan deviation of the raw data (shown in Supplementary Fig. 6a). g,h, O2 sensor response to changes in O2 concentration in distilled water at 37 °C before and after black silicone encapsulation (g) and ethylene oxide (EtO) sterilization (h). i, Data from the O2 sensors (n = 2); one sensor was incubated in PBS and the other sensor was incubated in undiluted human serum at 37 °C for 10 d. The 2-min O2 data from the sensors operated at 350 Hz fs were recorded at different times over 10 d. Data for 2 min were averaged and are presented as the mean ± 1 s.d.

The system response to various O2 concentrations and the Allan deviation of the data are illustrated in Fig. 4e,f. The system operated at 350 Hz fs exhibited an O2 sensitivity greater than ~0.066° (mm Hg)–1 and a phase (Ф) resolution less than 0.38° Hz–1/2, yielding an O2 resolution better than 5.8 mm Hg Hz–1/2 across the relevant O2 range of interest (<100 mm Hg). Because the system Ф resolution is dominated by jitter noise of the AFE and the luminescence intensity increases with a decreasing O2 concentration, the system SNR and hence Ф resolution improve for lower O2 levels.

The sensor O2 response is reversible (Fig. 4g). A calibration curve converts system phase output to O2 concentration (partial pressure of O2, pO2) (Supplementary Fig. 5). The sensor response times (the time required to reach 90% of the steady-state value) before black silicone encapsulation from ~3.5 to ~156 mm Hg O2 and vice versa were ~210 s and ~257 s, respectively; this increased to ~250 s and ~320 s after black silicone encapsulation.

The wireless sensors were sterilized in ethylene oxide, and their O2 response was tested to assess the potential effect of sterilization on sensor functionality. The sensor O2 response was nearly identical before and after sterilization (Fig. 4h), indicating that these sensors are sterilizable without loss of their functionality.

To evaluate the sensor’s ability to resist in vivo biofouling, we conducted in vitro experiments in PBS and undiluted human serum. Human serum, a complex fluid containing hundreds of different proteins29, was chosen to simulate in vivo fouling because non-specific protein adsorption (fouling) on the implant surface is considered the initial step in triggering a foreign body reaction30 and one of the critical factors causing the failure of many implants31. The sensors incubated in PBS and serum at 37 °C for 10 d demonstrated no apparent loss of O2 sensitivity (Fig. 4i).

The wireless system accurately reflects tissue oxygenation in a physiologically relevant large animal model

We next tested the clinical utility of the wireless direct O2-monitoring system in a physiologically relevant large animal model. Anesthetized juvenile or adult sheep (n = 2 animals) were intubated and mechanically ventilated. The biceps femoris was carefully dissected, and both the wireless sensor and a commercial wired pO2 sensor (NEOFOX) were placed below the muscle layer; the muscle and overlying skin were carefully closed (Fig. 5a–c). The US transducer, attached to a five-axis micromanipulator for fine alignment, was placed on top of the skin layer on an acoustic standoff pad (Fig. 5d).

Fig. 5: Surgical placement of and recording from the implantable, wireless O2 sensor under conditions of normoxia, hyperoxia and hypoxia.
figure 5

a, The wireless sensor was placed underneath the biceps femoris muscle in anesthetized sheep following careful surgical dissection. b,c, The biceps femoris muscle was layered over the sensor, completely covering it (b), and the overlying skin was sutured using an interrupted approach (c). d, The micromanipulator stage housing the US transducer was placed on top of the skin layer on an acoustic standoff pad. e, The anesthetized animal was provided with 100% inspired O2 (FiO2) via an endotracheal tube followed by a hypoxic gas mixture of 10% O2, achieved via nitrogen blending and confirmation with an inline O2 detector, followed by ventilation with room air (21% O2). Tissue O2 concentration readings were continuously monitored via the wireless O2 sensor and the wired commercial NEOFOX probe. Corresponding arterial pO2 (paO2), blood hemoglobin O2 saturation (SpO2) and FiO2 readings are provided. f, The stepwise reduction in FiO2 resulted in corresponding stepwise reductions in tissue pO2 readings that showed excellent concordance between the wireless O2 sensor and the commercial probe. Corresponding paO2, SpO2 and FiO2 readings are provided. Note that for FiO2 values above room air, SpO2 readings are unable to approximate paO2 or tissue pO2 levels.

As seen from the data (from animal A) in Fig. 5e and as determined by the wired pO2 sensor, acutely adjusting the inspired O2 concentration from 100% to 10% resulted in a rapid reduction in muscle pO2 from approximately 90 mm Hg to around 20 mm Hg. Similarly, the wireless sensor was able to accurately reflect muscle pO2 levels.

More gradual, stepwise reductions in FiO2 content resulted in more gradual reductions in tissue pO2, which were accurately determined by our wireless sensor in real time with similar kinetics to those of the commercial wired probe (Fig. 5f; data from animal B). Importantly, simultaneous determination of blood hemoglobin saturation via pulse oximetry was not capable of detecting differences in tissue or blood oxygenation above room air (21%), as all hemoglobin is fully saturated beyond this point.

In sum, our wireless O2 sensor system accurately reflects tissue oxygenation status under physiological states and therefore has potential to augment clinical decision making in settings where tissue or an individual’s oxygenation status warrants careful monitoring.

The system demonstrates robust wireless operation both in vitro and ex vivo with the implant operated at a 10-cm depth

To further evaluate uplink performance, the system was operated at 350 Hz fs and a 10-cm depth in distilled water with and without intentional misalignment and through fresh, ex vivo porcine tissue (which presents inhomogeneous acoustic properties) (Fig. 6). The system, while operating without any misalignment and using a 2-MHz acoustic field with an ISPTA of 282 mW cm–2, exhibited an MD of over 14%, a BER < 10−5 and an US link PTE of ~0.74%. When operated with an intentional transverse offset of ~1.21 mm between the central axes of the acoustic field and the sensor piezo, the system demonstrated an MD of ~108% and a BER < 10−5, but at the expense of a ~58.9% increase in ISPTA, resulting in a concomitant reduction in the US link PTE. The MD increase was mainly due to reduced reflections from non-responsive regions of the sensor, indicating that an acoustic field with a smaller transverse beam-spot size (that is, comparable to the piezo size) can provide more robust uplink performance and higher US link PTE.

Fig. 6: In vitro and ex vivo uplink characterization of the wireless O2-sensing system.
figure 6

a,d, Wireless sensor operation at a 10-cm depth in distilled water without (a) and with (d) an intentional transverse offset of the central axis of the acoustic field to that of the sensor’s piezo. b,e, Sensor waveform and backscatter signal were captured in the wireless measurement of a single O2 sample for the sensor operated in vitro at a 10-cm depth without (b) and with (e) a transverse misalignment. c,f, Backscatter relative difference for wirelessly recorded O2 samples (≥100,000) from the sensor operated in vitro at a 10-cm depth without (c) and with (f) a transverse misalignment. g, The O2 sensor was wirelessly operated at a 10-cm depth through an inhomogeneous sample of fresh, ex vivo porcine tissue, in which acoustic waves passed through approximately 3 mm of US gel, 1 mm of skin, 1 mm of fat and 95 mm of muscle tissue; scale bar, 1 cm. h, Sensor waveform and backscatter signal were recorded in the wireless measurement of a single O2 sample for the sensor operated at a 10-cm depth through a porcine specimen. i, Backscatter relative difference for 100,000 O2 samples, showing ~21% MD from the sensor operated at a 10-cm depth through a porcine tissue specimen. The system achieved an uplink BER < 10−5 in the measurement.

In the ex vivo measurement, the external transducer generated US waves with 706 mW cm–2 ISPTA derated by 0.3 dB cm−1 MHz−1 (FDA standard US attenuation in soft tissue13), producing ~228 mW of acoustic power at the transducer surface that propagated through approximately 3 mm of US gel and 97 mm of tissue. The system achieved a ~21% MD and a BER < 10−5, demonstrating a robust uplink, and a US link PTE of ~0.08% with the sensor consuming ~194 μW of average electrical power. Note that system uplink performance (MD and BER) depends on acoustic attenuation in heterogeneous tissue, which varies for different tissue types and specimens. This is because acoustic absorption and dispersion may change the amplitude of the unmodulated backscatter signal received by the external transceiver during the time interval within which the uplink data are received. Furthermore, US link PTE also depends on acoustic attenuation; for example, the system operated through a porcine tissue specimen exhibited a significantly lower link PTE than the system operated in water because the tissue sample has much higher US attenuation than that of water.

Discussion

This work presents the miniaturization of a fully implantable optrode in the cubic millimeter volume range, which is suitable for deep-tissue measurements. Although the focus of the study is a system for measuring O2 tension in vivo, the fundamental technological achievements open the door to minimally invasive pulse oximetric sensors, pH sensors and CO2 sensors, among others. Each of these, embodied in ultraminiature deep-tissue systems, would enable new diagnostics.

Regarding the measurement of deep-tissue O2 tension, the application space of the wireless O2-sensing system is vast. Organ transplantation provides a clear example. Monitoring graft oxygenation following organ transplantation is critical, but relies on indirect methods that require skilled operators and provide only intermittent snapshots of tissue perfusion32,33. Continuous and reliable monitoring of graft oxygenation following orthotopic liver transplantation, for example, may enable the early detection of graft ischemia due to either hepatic artery thrombosis or graft vascular disease, allowing for timely surgical re-exploration to minimize the risk of graft loss, which could be fatal2,34. Importantly, these complications can occur months to years after transplantation35,36. Furthermore, more than 5.7 million individuals are admitted annually to intensive care units in the United States37,38. Assessment of tissue oxygenation is a fundamental need in this setting. Although the current iteration of our device requires surgical placement, future miniaturization efforts may enable semi-invasive/vascular approaches for probe placement. Global measures of cardiopulmonary performance, such as cardiac output, O2 delivery or blood pressure, frequently do not reflect local metabolic demands at the organ and tissue levels and can promote excessive fluid loading or inotrope dosing, worsening outcomes39,40,41. An important contributor in this setting is a lack of hemodynamic coherence between the microcirculation and the macrocirculation42. Coupling direct measurements of the microcirculation43 with direct monitoring of tissue pO2 would greatly augment critical care management approaches.

To clinically adopt the wireless O2-sensing system for the use of chronic in vivo O2 tracking, a number of technical challenges must be addressed. The first challenge is implant localization by an external transceiver, because the postsurgical in vivo position may drift relative to any external fiducials; such movement can arise due to pressure from outside of the body, movement or breathing, and scar formation. The in vivo localization before each pO2 measurement can be achieved with a phased-array transceiver that uses US backscatter information first to find and then to track the time-dependent implant position in the body44.

A second challenge arises because acoustic attenuation due to absorption and scattering varies between different US propagation paths in heterogeneous tissue; a path with higher attenuation may significantly degrade system power transfer efficiency and data transfer reliability. For example, muscle tissue with a more unevenly distributed intramuscular fat content will exhibit greater acoustic attenuation45. Here, an external transceiver with a large-aperture, multi-element transducer array capable of focusing energy to the implant will allow steering of the US beam along a preferred path. Finally, a phased array could also potentially be used to interrogate multiple O2 sensors implanted in different locations of target tissue, in a time-division multiplexing fashion or simultaneously46.

Lastly, chronic in vivo use of the wireless O2 sensor will require hermetic packaging to prevent biofluid penetration into the electronic sensor components and the piezocrystal. Traditionally, small implantable housings make use of ceramic or titanium enclosures brazed or microwelded to achieve the required hermeticity. This is an active area of work both commercially and academically; an extensive review was recently published47. Acoustic windows for efficient ultrasonic energy transfer into ceramic or metallic housings have recently been demonstrated in the literature48. For this work, we used biocompatible polymers to encapsulate the implant, given their ease of use for acute and semichronic experiments. It is well known that polymeric materials at these thicknesses are not suitable for long-term in vivo use due to their high water vapor permeability49.

Methods

Fabrication of the O2-sensing film

Film fabrication included two steps. First, a luminescent dye, tris-(bathophenanthroline) ruthenium (II) perchlorate (Ru(dpp)3(ClO4)2) (75213-31-9, GFS Chemicals), was immobilized on the surface of silica particles (LiChrosorb Si 100 (10 µm); 7631-86-9, Sigma-Aldrich) at a ~1:10 dye:particle ratio by weight. Briefly, 200 mg of Ru(dpp)3(ClO4)2 complex was dissolved in 10 ml of ethanol (ACS reagent ≥99.5%; 459844, Sigma-Aldrich). Silica gel was prepared by adding 2 g of silica particles to 40 ml of aqueous sodium hydroxide solution (0.01N; 1310-73-2, Thermo Fisher Scientific) and magnetically stirring the mixture at 1,000 r.p.m. for 30 min. Next, the dye-containing ethanol solution was poured into the silica gel solution and stirred at 1,000 r.p.m. for 30 min. The dye-containing silica particles were filtered out of the solution through a filter with a pore size of 0.45 µm (165-0045, Thermo Fisher Scientific), washed once in ethanol and three times in deionized water. All of the supernatant was removed, and the dye-loaded silica particles were dried at 70 °C overnight.

Second, the dye-loaded silica particles were incorporated into PDMS to avoid problems related to dye leaching in aqueous media. Note that PDMS was selected as the polymer matrix due to its excellent O2 permeability, biocompatibility, optical transparency and high solubility for the fillers, preventing agglomeration50. Dried silica particles (2 g) were thoroughly mixed with 20 g of PDMS prepolymer part A and 2 g of PDMS curing agent part B (Sylgard 184, Dow Corning). A ~100-µm-thick film was prepared by spinning a small amount of this mixture at 500 r.p.m. on a microscope slide and curing it at 60 °C under vacuum (<10 torr) in the dark for ~7 d to remove the solvent and air bubbles. The cured film was kept in the dark at room temperature for at least 24 h before use and was stored in the dark at room temperature. Note that the amount of silica particles in PDMS and the film thickness were adjusted to maintain a reasonable trade-off between luminescence intensity and O2 response time23,51

Design, fabrication and assembly of the wireless sensor

The sensor (implant) was built on a 100-µm-thick polyimide, flexible PCB with electroless nickel/immersion gold coating (Rigiflex Technology). A 750-µm-thick PZT sheet with silver electrodes was diced using a dicing saw with a 300-µm-thick ceramic-cutting blade. A PZT cube (750 µm3) was first attached to a PCB using two-part conductive silver epoxy with a 1:1 mix ratio (8331, MG Chemicals), and the board was cured at 65 °C for 15 min, well below the PZT Curie temperature and melting temperature of polyimide. The PZT top electrode was wire bonded to the PCB using a wedge bonder (747677E, West Bond) to create an electrical connection between the PZT and IC. The board was encapsulated with ~10-μm-thick parylene-C using chemical vapor deposition (Specialty Coating Systems) to insulate the piezocrystal from the environment due to its biological inertness and resistance to moisture52. For the IC and its wire bonds, the PCB’s open metal pads were carefully exposed by scoring the parylene around the pads using a sharp probe tip and removing the parylene layer. The IC was attached to the PCB using the same silver epoxy, cured at 65 °C for 15 min and wire bonded to the PCB. Next, a ~250-µm-thick optical filter (Edmund Optics) was attached to the top of the IC using medical-grade UV-curable epoxy (OG142, Epotek). The same UV-curable epoxy was also used to assemble other sensor components, including a 650 × 350 × 200 µm3 µLED (APG0603PBC, Kingbright) and its 3D-printed holder (Protolabs), and to protect and insulate the wire bonds of the chip and µLED. After assembly, the O2-sensing film was slipped through the gap between the µLED holder and the optical filter. The small residual space between the µLED holder and the film was filled with PDMS (Sylgard 184, Dow Corning). PDMS parts A and B were mixed in a 10:1 ratio, degassed, poured between the space and cured at room temperature for 48 h. Finally, the O2-sensing region on the sensor surface was coated with a ~180-µm-thick layer of biocompatible black silicone. The black silicone consisted of two-part, low-viscosity silicone elastomer (MED4-4220, NuSil Technology) and black, single-component masterbatch (Med-4900-2, NuSil Technology); the two silicone parts were first mixed in a 1:1 weight ratio, and the masterbatch (4% by weight) was added, thoroughly mixed, degassed for <~5 min, applied and cured at room temperature for 48 h.

Here, PZT was selected as a piezoelectric material due to its high electromechanical coupling coefficient and high mechanical quality factor, providing high power harvesting efficiency. A lead-free biocompatible barium titanate (BaTiO3) ceramic with a slightly lower electromechanical coupling coefficient can be used in place of PZT53.

The detection volume of the O2 sensor was estimated from the material volume where O2 molecules diffuse through to the O2-sensing film under the µLED. The implant volume was measured by suspending the implant without test leads with a thin, rigid wire below the water surface in a container placed on an electronic balance with 0.1-mg measurement accuracy. The implant volume was calculated from the weight difference of a water-filled container before and after submersion of the implant in water; the weight difference, equal to the buoyant force, was divided by the water density to determine the actual implant volume. Measurements were performed using two implants; each implant volume was measured five times to determine reproducibility. The obtained data were presented as the mean ± 2 s.d.

Optical characterization

The absorption spectrum of the O2-sensing film and the transmission spectrum of the optical filter were measured with a spectrophotometer (Jenway 6300). The emission spectra of the film and µLED were measured using a CCD spectrometer (Thorlabs, CCS200/M) operating at a 1-s integration time and enabled with electric dark correction. The film samples were excited at 450 nm by a laser diode (Osram, PL450B) driven with a Keithley 2400 source meter, and its emission was scanned in the range of 515–800 nm. The µLED optical output power was measured using an optical power meter (Thorlabs, PM100D) equipped with an Si photodiode detector (Thorlabs, S121C). The current–voltage curve of the µLED was measured with a Keithley 2400 source meter. Photodiode responsivity was characterized using a halogen lamp coupled to a monochromator, a reference photodiode (Thorlabs, FD11A) and an Agilent B2912A source meter. The same photodiode (FD11A) was also used to measure the µLED output light intensity.

Photobleaching and leaching tests

Dye photobleaching was evaluated using a fully packaged sensor continuously operated in room air (21% O2) at room temperature. After the bleaching test, the same sensor was continuously operated in 1× PBS without calcium and magnesium (Corning, Mediatech) at 37 ± 0.1 °C in an oven (TestEquity Model 107) to assess dye leaching and further photobleaching. In the tests, the sensor was driven by differential 2-MHz AC signals from a Keysight 33500B function generator, which are ac-coupled to the IC’s rectifier inputs. The transimpedance amplifier (TIA) output (Supplementary Fig. 3) was connected to a buffer (LTC6268, Linear Technology). The buffer output was continuously measured using a 16-bit digitizer (NI myDAQ; National Instruments) with 200-kHz sampling rate. A custom Labview program (Labview 2018, National Instruments) was developed to detect and record the peak-to-peak amplitude of the buffer output that is directly proportional to the Ru-dye luminesce intensity.

Design of the external transceiver

The external transceiver consisted of TX and RX paths. The TX path included a high-voltage pulser with an integrated TX/RX switch (MAX14808, Maxim Integrated) and a digital controller module (NI PXIe-6363, National Instruments). During the TX mode, the high-voltage pulser converted a low-voltage signal from the digital controller module to a high-voltage signal to drive an external transducer. The RX path included an ultralow noise amplifier (AD8432, Analog Devices) to receive and amplify the backscatter signal from the external transducer, a gain amplifier to further amplify the signal to a level within the input range of the analog–digital converter (ADC) and a digitizer with an antialiasing filter and a 14-bit high-speed ADC (NI PXIe-5122, National Instruments) to filter and digitize the signal after receiving and amplification. In addition to the switch integrated into the pulser, an additional switch (ADG619, Analog Devices) was used to minimize the electrical coupling between the TX and RX paths. TX and RX paths were synchronized using the clock integrated into the backplane of the PXI chassis (NI PXIe-1062Q, National Instruments). The digital controller, digitizer and NI PXIe-8360 modules were inserted in the chassis, in which the NI PXIe-8360 module was used to connect the chassis to a computer for communication with the other modules and data transfer.

A custom Labview program was developed to control the modules and to process the backscatter data in real time. A (TX and RX) communication protocol was encoded in the program. During data processing, the backscatter data digitized by a 14-bit ADC with a 20-MHz sampling rate were resampled by a factor of five and then interpolated with a sinc function. The sinc interpolation was followed by peak detection to extract the envelope of the backscatter signal and linear interpolation to increase the number of data points, improving the accuracy in determining an optimal threshold that minimizes BER. An optimal threshold (that is, the half value of the sum of modulated and unmodulated backscatter signal amplitudes) was determined by taking the mean of the data points from the time intervals where the steady-state backscatter signal was amplitude modulated and unmodulated. The threshold was used to convert the digitized data into digital format (bits, ‘0’ or ‘1’). The bits were scanned to find a preamble and a postamble and hence to extract O2 data bits. The binary-coded data were converted to numeric data, which were sent to a computer through a serial link for data storage and further analysis.

In vitro and ex vivo characterization

An in vitro characterization of the system (Fig. 4 and Supplementary Figs. 7 and 8) was performed in a custom-built water tank (Supplementary Fig. 9) using a 25.4-mm diameter, 2.25-MHz single-element ultrasonic transducer with a 47.8-mm focal depth (V304-SU-F1.88IN-PTF, Olympus). During measurements, the transducer face was covered with a thin latex sheet by filling the empty space between the transducer face and the latex sheet with castor oil (acting as a coupling medium) to protect the transducer-matching layer from possible damage due to long-term contact with water. A hydrophone (HGL-0400, Onda) was used to calibrate the output pressure and characterize the acoustic beam patterns of the external transducer (see Supplementary Fig. 10 for the beam patterns of the transducers used in this study). An Allan deviation analysis54 was used to quantify the system noise performance (Fig. 4).

During measurements, the water tank was placed on a stirring hotplate (Thermo Scientific Cimarec) to keep the water temperature constant at 37 ± 0.1 °C and to stir the water to increase the speed of transition from a low to high O2 level in distilled water and vice versa. The water O2 concentration was monitored using a commercial O2 probe with a 300-µm core diameter (NEOFOX-KIT-PROBE; BIFBORO-300-2; Ocean Optics) and was varied by controlling the ratio of O2 and N2, which were supplied to the water tank through two pipes via two identical gas flow controllers (FMA-A2407, Omega) connected to O2 and N2 gas cylinders. A custom Matlab program controlled the flow controllers through a digital–analog converter (NI myDAQ, National instruments).

Ethylene oxide sterilization with an exposure time of 4 h at 37 ± 3 °C and an aeration time of 24 h at 37 ± 3 °C was performed by a commercial vendor (Blue Line Sterilization Services).

To assess the functionality of the wireless O2 sensors over time in an environment that mimics (to first order) in vivo biofouling, sensors were incubated with 1× PBS without calcium and magnesium (Corning, Mediatech) and pooled human serum (off the clot) (Innovative Research) for 10 d at 37 ± 0.1 °C in an oven (TestEquity Model 107). Two antibiotics, penicillin (100 U ml–1) and streptomycin (100 μg ml–1) (Gibco by Life Technologies, 15-140-122; Thermo Fisher Scientific), were added to human serum to inhibit bacterial growth. The test in serum was performed by placing the sensor in a container, and antibiotic-supplemented serum was replaced every 24 h to ensure sterile conditions during the study. During testing, the sensors were operated at a 350-Hz sampling rate and differential 2-MHz AC signals were produced by a Keysight 33500B function generator, which are ac-coupled to the IC’s rectifier inputs. One of the rectifier inputs was connected to a high-input impedance buffer amplifier (LTC6268, Linear Technology). The buffer output, O2 data, was recorded by a 14-bit digitizer (NI PXIe-5122, National Instruments) synchronized to the function generator and a custom Labview program.

Measurements for system uplink performance assessment at a 10-cm depth in distilled water and in a fresh ex vivo porcine specimen were performed with a custom, spherically focused 2-MHz, 25.4-mm diameter ultrasonic transducer with an 88.1-mm focal length (Sensor Networks; Fig. 6 and Supplementary Fig. 11a). An additional measurement for system uplink performance assessment at a 5-cm depth in a fresh ex vivo porcine specimen was performed using a 25.4-mm diameter, 2.25-MHz single-element ultrasonic transducer with a 47.8-mm focal depth (V304-SU-F1.88IN-PTF, Olympus; Supplementary Figs. 11b and 12). During measurement, a porcine tissue sample was positioned between the wireless sensor and the external transducer, with coupling enabled by US gel (Aquasonic Clear; Parker Labs). Air bubbles in the US gel were removed via centrifugation at 2,800 r.p.m. for 10 min. To remove possible air bubbles entrapped between the sensor and the tissue, the sensor was positioned on a tissue sample in a container filled with distilled water. An US absorber was placed under the tissue sample to avoid US reflection from the container bottom interface.

Backscatter modulation depth

Backscatter relative difference (BRD) is defined as the ratio of the amplitude difference between the modulated and unmodulated backscatter signals to the amplitude of the unmodulated backscatter signal. The MD percentage was calculated by multiplying the BRD by 100. BRD plots were obtained by collecting data samples from time points where the steady-state backscatter signal was amplitude modulated and unmodulated during O2 measurement.

In vivo measurements

The Committee on Animal Research of the University of California, San Francisco, approved all protocols and procedures. All animal research conforms to the National Institutes of Health (NIH) guidelines as outlined in the Guide for the Care and Use of Laboratory Animals.

Lambs (2–12 weeks of age) were given ketamine (10–15 mg kg–1, intramuscularly), and an appropriately sized catheter was placed percutaneously into a peripheral vein. The lambs were given intravenous (IV) boluses of fentanyl citrate and diazepam for sedation and anesthetized with continuous IV infusions of fentanyl citrate, ketamine hydrochloride and diazepam. Heart rate and blood pressure were continuously monitored, and doses were adjusted with any changes in these parameters. Catheters were placed in an artery and vein of each hind limb. This procedure was done either before or after endotracheal intubation when a surgical plane of anesthesia was present. Lambs were intubated with a 4.5–7.5-mm OD cuffed endotracheal tube and mechanically ventilated with a pediatric time-cycled, pressure-limited ventilator. Vecuronium bromide was given intermittently (IV bolus) for muscle relaxation and redosed for movement. Ventilation with 21% O2 was adjusted to maintain a systemic arterial pCO2 between 35 and 45 torr. To ensure that lambs were adequately anesthetized during paralysis, heart rate, systemic arterial pressure, temperature and response to stimuli were continuously monitored. In addition, arterial blood gas readings were performed every 30–120 min to determine pCO2, pO2 and pH parameters.

Following instrumentation, animals were allowed to recover to steady state until they required no further adjustment to sedatives and exhibited stable hemodynamic parameters. This time was designated as the normoxic baseline, and blood gas analysis was performed to confirm physiologic parameters.

To facilitate O2 sensor placement, a two-way skin flap of an appropriate size was created over the lateral thigh. The uppermost layers of skin were carefully dissected away from the subcutaneous space to ensure tissue integrity and were then reflected away from the site. Exposure of the posterior fascial compartment was then visualized and dissected away to reveal the biceps femoris muscle. Because of muscle depth, the lateral aspect of the biceps femoris was carefully dissected to create a tunnel into the underlying fascial plane. The wireless O2 sensor was then seated under the biceps femoris via the created tunnel, and the skin was closed in a simple continuous fashion. The wireless O2 sensor had five flexible leads used only for electrical testing and debugging before in vivo measurements, and, as seen in Fig. 5, they were not accessible from the outside except two leads (ground and VDC–DC) in the in vivo measurements, which were used to check the sensor functionality after implantation and were completely disconnected during measurements. A wired O2 probe (NEOFOX-KIT-PROBE; BIFBORO-300-2; Ocean Optics) was placed nearby to confirm tissue pO2 measurements. Tissue pO2 measurements were compared to paO2 obtained by blood gas analysis as well as with SpO2 obtained via a GE Dash 3000 patient monitor and GE Masimo finger probe placed on the ear.

At the end of the protocol, all lambs were euthanized with a lethal injection of sodium pentobarbital (150 mg kg–1) followed by bilateral thoracotomy, as described in the National Institutes of Health Guidelines for the Care and Use of Laboratory Animals.

In vivo experiments were performed using a 12.7-mm diameter, 2.25-MHz ultrasonic transducer with a 21.6-mm focal depth (V306-SU-F0.88IN-PTF; Olympus), which demonstrated similar beam patterns with the 25.4-mm diameter transducer used in the in vitro experiments (Supplementary Fig. 10). The transducer output pressure and acoustic intensity were characterized in distilled water using a hydrophone (HGL-0400; Onda) after its face was covered with castor oil and a thin latex sheet to protect its matching layer from possible damage due to long-term direct contact with US gel. The external transducer was attached to a custom five-axis micromanipulator built using single-axis linear stages (DT-100 X; SD Instruments), a tilting stage (MM-2A; Newport) and a 0.5-mm acrylic sheet. The micromanipulator stage, used for alignment of the external transducer to the implanted sensor, was placed on top of the skin layer on a 7-mm-thick acoustic standoff pad (DIY ultrasound phantom gel; Humimic Medical) with US gel (Aquasonic Clear; Parker Labs) applied to the skin surface and the external transducer for acoustic coupling. The US backscatter amplitude (Supplementary Fig. 13) was used for fine alignment of the implant and the external transducer.

Tissue pO2 measurements were performed with the wireless system operated at a sampling rate of 350 samples per second. In the in vivo measurements (Fig. 5 and Supplementary Fig. 11c–e), the maximum distance from the external transducer to the wireless O2 sensor, operated with an acoustic field that had a derated ISPTA of 454 mW cm–2 and a mechanical index of 0.08 (both below the FDA regulatory limits of 720 mW cm–2 and 0.19, respectively), was ~26 mm, with ~19 mm consisting of tissue (including skin, fat and muscle). The distance between the implanted sensor and the external transducer was estimated from the 2ToF (that is, the time delay between the received backscatter signal from the sensor piezo and the signal that drove the external transducer). The data from the wireless O2 sensor and the wired O2 probe, shown in Fig. 5, were averaged every 5 s. Two identical wireless O2 sensors were used in the in vivo measurements; the first sensor O2 responses in water and animal A are shown in Figs. 4e and 5e, respectively, and the second sensor O2 responses in water and animal B are shown in Supplementary Fig. 14 and Fig. 5f. All images were captured by a smartphone camera.